Sequencing Knowledge Base

Library Structure

  • Illumina Adapter Sequences: a comprehensive list of all Illumina adapters and indices sequences from various products (including nextera, truseq, etc.)
  • Standard Index oligos:
    • P7: 5’ CAAGCAGAAGACGGCATACGAGAT[index][adapter_sequence]3’
    • P5: 5′ AATGATACGGCGACCACCGA[dual_index][adapter_sequence] 3′
      Screen Shot 2020-07-30 at 12.37.32 AM

Laser and Dyes

  • Nextseq and Novaseq – 2 dye
    • Rather than a separate dye for each base, 2-channel SBS uses a mix of dyes. Images are taken of each DNA cluster using red and green wavelength filter bands.
    • Clusters seen in red or green images are interpreted as C and T bases, respectively. Clusters observed in both red and green images are flagged as A bases (appearing as yellow clusters), while unlabeled clusters are identified as G bases).                               
         laser-1
  • Miseq – 4 dyes/2 lasers
    • G/T share one laser and C/A share another. When choosing indexes, make sure each base has diversity between the two lasers.

laser2

                 laser3

  • iSeq
    • One dye and two chemistry step
               iseq
  • For all sequencers: there is no index primer site for i7 in PhiX therefore spiking in more phix would not help with index diversity. Multiplexing under six samples requires careful selection of barcodes. 

Illumina Sequencing Outputs

  • For all sequencers, the minimum R1 cycle is 25 (26 for NovaSeq)
Illumina Kit Cluster Density (k/mm2) Reads
NextSeq 500/550 High Output v2 (75 cycle) 170-220 400M
NextSeq 500/550 High Output v2 (150 cycle) 170-220 400M
NextSeq 500/550 High Output v2 (300 cycle) 170-220 400M
NextSeq 500/550 Mid Output v2 (150 cycle) 170-220 130M
NextSeq 500/550 Mid Output v2 (300 cycle) 170-220 130M
NextSeq 2000 P3 (100 cycle) N/A 1.1B
NextSeq 2000 P3 (300 cycle) N/A 1.1B
MiSeq Kit v3 (150 cycle) 1200-1400 25M
MiSeq Kit v3 (600 cycle) 1200-1400 25M
MiSeq Kit v2 (50 cycle) 1000-1200 15M
MiSeq Kit v2 (300 cycle) 1000-1200 15M
MiSeq Kit v2 (500 cycle) 1000-1200 15M
MiSeq Micro Kit v2 (300 cycle) 1000-1200 4M
MiSeq Nano Kit v2 (300 cycle) 1000-1200 1M
MiSeq Nano Kit v2 (500 cycle) 1000-1200 1M
Novaseq S4 (300 cycle) N/A 8-10B
Novaseq S2 (100 cycle) N/A 2.8 – 3.3B
Novaseq S2 (200 cycle) N/A 2.8 – 3.3B
Novaseq S2 (300 cycle) N/A 2.8 – 3.3B
Novaseq SP (500 cycle) N/A 800M

Maximum cycles in a SBS kit

  • There is overage for extra cycles built into every kit. Operators could use each to the fullest extent by calculating the maximum cycles for read 1 and read 2. Please adjust accordingly, as some chemistry need additional cycles for dual index read (also known as dark cycling).
    • Example: A NovaSeq S2 300 cycle kit has 325 cycles, 325 – 8 (index1) – 8 (index2) – 7 (dark cycles) = 302 cycles remaining for read 1 and read 2. This kit is validated for up to a paired end 151 cycles.

Screen Shot 2020-07-30 at 12.37.21 AM

How many reads do I need?